19510501-V2.doc 15.07.13 Operating Manual ME English Heating Immersion Circulator ME 1.951.0501-V2 06/13

Pagina 2 - Unpacking and inspecting

Operator responsibility – Safety recommendations 10 Safety installations according to IEC 61010-2-010: Excess temperature protection adjustable fro

Pagina 3

ME 11 Operating instructions 3. Safety notes for the user 3.1. Explanation of safety notes In addition to the safety warnings listed, warnings are

Pagina 4

Safety notes for the user 12 3.3. Safety recommendations Follow the safety instructions to avoid personal injury and property damage. Also, the val

Pagina 5 - 1.1. Description

ME 13 Caution: The temperature controlling i.e. of fluids in a reactor constitutes normal circulator practice. We do not know which substances are

Pagina 6 - Appropriate operation

Operating controls and functional elements 14 4. Operating controls and functional elements 4.1. Circulator Front view Rear view 1 Mains p

Pagina 7 - 2.1. Disposal

ME 15 Menu keys 5 Key: start the menu > warning and safety values< 6 Key: start the menu >temperature setpoints< 7 Key: display of

Pagina 8 - 2.3. Warranty conditions

Preparations 16 5. Preparations 5.1. Installation Caution: Securely fix the immersion circulator. The heater may not be in contact with the wall of t

Pagina 9 - 115 V/60 Hz

ME 17 Recommended bath fluids: Bath fluid Temperature range soft/decalcified water 5 °C to 80 °C JULABO bath fluids JULABO Description T

Pagina 10

Preparations 18 5.3. Temperature application to external systems Caution: Securely attach all tubing to prevent slipping. If the circulator is oper

Pagina 11 - Operating instructions

ME 19 *) Adapter for metal tubing M10x1 on M16x1 Order No. 8 970 444 Warning: Tubing: At high working temperatures the tubing used for temp

Pagina 12 - 3.3. Safety recommendations

2 Congratulations! You have made an excellent choice. JULABO thanks you for the trust you have placed in us. This operating manual has been design

Pagina 13

Preparations 20 Pt100 RS 23215AMP15AMPVJULABO MVS2725a 25b26  For applications near ambient temperature (20 °C), the cooling water temperature shoul

Pagina 14 - 4.1. Circulator

ME 21 6. Operating procedures 6.1. Power connection Caution:  Only connect the unit to a power socket with earthing contact (PE – protective earth)

Pagina 15 - Remote control

Setting of temperatures 22 7. Setting of temperatures 7.1. Using the pre-settings in the menu Press the key to call up the menu for temperatu

Pagina 16 - 5. Preparations

ME 23 7.2. Direct setting of temperatures The circulator uses the setpoint of SETPNT 1 or 2 or 3 for temperature control. The indicated setpoi

Pagina 17

Safety installations, warning functions 24 8. Safety installations, warning functions Check the safety installations at least twice a year! Refer

Pagina 18 - Recommended tubing:

ME 25 8.1.1. Early warning system, low level protection This low level protection is independent of the control circuit and is divided into two

Pagina 19 - 5.2. Countercooling

Safety installations, warning functions 26 8.2. Switch-over from warning to shutdown function If a shutdown of functional elements (e.g. heater, ci

Pagina 20 - JULABO MVS

ME 27 8.3. Over and Sub temperature warning function Over temperature Sub temperature If the observance of a working temperature value >SE

Pagina 21 - 6. Operating procedures

Menu functions 28 9. Menu functions  Menu level 1 The term „Menu functions“ refers to settings such as  Start program Page 29 Administration

Pagina 22 - 7. Setting of temperatures

ME 29 9.1. MENU PROGRAM – START Start-Menu This menu will start a previously set program.  Requirements: 1. Create a program. (refer to next c

Pagina 23

ME 3 TABLE OF CONTENTS Operating manual ...

Pagina 24

Menu functions 30  Submenu TIMER Parameter level  set the time for the start of the program in the submenu >TIMER<.  >TIME<

Pagina 25

ME 31 C2 D1 D2 E xxx.xx – external actual value D RUN – the program has started or PAUSE – the progress of the program has been interrupted by pr

Pagina 26 -  Setting >WARNING<

Menu functions 32 Menu level 1  >EDIT< Create, administer program > STEP< Program step (1 ... 10) >SETPNT < Temperature

Pagina 27

ME 33 9.3. MENU PUMP – Setting of pump pressure Factory setting: stage 2 The capacity of the circulating pump is set by adjusting the motor sp

Pagina 28 - 9. Menu functions

Menu functions 34 9.4. MENU CONFIG – Configuration of unit Menu level 1   A RESET can be effected only in the >OFF< mode. Switch off c

Pagina 29 -  Requirements:

ME 35 Level 3 Parameter level   Year flashes, set by pressing +   Return to factory settings by pressing RESET returns all set

Pagina 30

Menu functions 36 interface. If such a safety standard is not required, the NAMUR recommendations can be bypassed with the AUTOSTART function thus all

Pagina 31 - Step 1 2 4 6 8

ME 37 9.5. MENU CONTROL – Control characteristics and parameters Menu level 1  The circulator is qualified for internal and external

Pagina 32

Menu functions 38 Level 2 Parameter level  0 … 999  The parameter flashes, set by pressing +  0.1 … 99.9  The parameter flashes,

Pagina 33

ME 39 9.5.2. Dynamic internal °CtAPERNORMSetpointtemp. stability °CtAPERNORMtemperature ramptemp. stabilityThis parameter affects the temperature se

Pagina 34

4 8.3. Over and Sub temperature warning function ... 27 9. Menu functions ...

Pagina 35 - R“ for remote

Menu functions 40 9.5.4. Control parameters– XP-, TN-, TV- INTERNAL In most cases the control parameters preset in the factory are adequate for ach

Pagina 36 - 9.4.3. OFF-MODE

ME 41 9.6. MENU SERIAL - BAUDRATE, HANDSHAKE, PARITY Menu level 1  For communication between circulator and a PC or a superordinated process con

Pagina 37 - C-TYPE INTERNAL

Menu functions 42 9.7. MENU ATC - Absolut Temperature Calibration ATC serves to compensate a temperature difference that might occur between circul

Pagina 38

ME 43 Menu level 1  Level 2 Parameter level  Press the key if parameter is to be retained. Correction function for parameters or values (pri

Pagina 39 - 9.5.2. Dynamic internal

Menu functions 44  The value is only indicated   Integer digits flash, set by pressing +  Decimal digits flash, set by pressing +

Pagina 40

ME 45 9.7.3. CALIBRATION TYPE: 1 -/ 2 -/ 3 POINT A >1-point<, >2-point< or >3-point< calibration can be carried out. First geo

Pagina 41

Menu functions 46 9.7.4. Example: 3-point calibration for internal control In the temperature range from 80 °C to 160 °C the calibration curve of

Pagina 42 - M B (INT) T

ME 47   The parameter flashes, switch by pressing and . A >3-point< calibration is carried out.  The value >TMPVAL< is on

Pagina 43

Menu functions 48 9.8. MENU LIMITS Menu level 1  Level 2 Parameter level Press the key if parameter is to be retained. Correction function

Pagina 44

ME 49 9.8.1. Limits for internal control SETPOINT MAX / MIN – Maximum and minimum setpoint Restriction of the adjustable temperature range. Th

Pagina 45 - Pairs of values:

ME 5 Operating manual 1. Intended use JULABO circulators have been designed to control the temperature of specific fluids in a bath tank. JULABO

Pagina 46

Menu functions 50 BAND HIGH / LOW – Band limitation The band limitation is active during external control. Varied, practice-oriented setting are fe

Pagina 47

ME 51 10. Troubleshooting guide / error messages + Alarm with complete shutdown: If one of the following failures occur a complete, all-pole shut

Pagina 48 - 9.8. MENU LIMITS

Troubleshooting guide / error messages 52 Error message with ticker: >SENSOR DIFFERENCE ALARM-CHECK VISCOSITY AND PUMP STAGE< Defect of working

Pagina 49 -  SET MAX > SET MIN

ME 53 11. Electrical connections Notice: Use shielded cables only. The shield of the connecting cable is electrically connected to the plug housing.

Pagina 50 - BAND LOW

Remote control 54 / Control output The connector may be used for control of JULABO refrigerated circulators or as output for alarm messages. 1324

Pagina 51 - Alarm without shutdown:

ME 55 12.2. Communication with a PC or a superordinated data system If the circulator is put into remote control mode via the configuration level,

Pagina 52

Remote control 56 12.3. List of commands out commands: Setting temperature values or parameters. Command Parameter Response of circulator out_mode

Pagina 53 - 11. Electrical connections

ME 57 in commands: Asking for parameters or temperature values to be displayed. Command Parameter Response of circulator version none Number of s

Pagina 54 - 12. Remote control

Remote control 58 Command Parameter Response of circulator in_mode_01 none Selected setpoint: 0 = SETPNT 1 1 = SETPNT 2 2 = SETPNT 3 3 = Last setpo

Pagina 55 -  end of file (; Hex: 0D)

ME 59 Error messages Description -06 SENSOR DIFFERENCE ALARM Sensor difference alarm. Working temperature and safety sensors report a temperature dif

Pagina 56 - 12.3. List of commands

Operator responsibility – Safety recommendations 6 2. Operator responsibility – Safety recommendations The products of JULABO ensure safe operation

Pagina 57

JULABO Service – Online remote diagnosis 60 13. JULABO Service – Online remote diagnosis JULABO circulators of the HighTech series are equipped w

Pagina 58

ME 61 14. Cleaning / repairing the unit Caution:  Always turn off the unit and disconnect the mains cable from the power source before cleaning

Pagina 59

ME 7 Use: The bath can be filled with flammable materials. Fire hazard! There might be chemical dangers depending on the bath medium used. Observe all

Pagina 60

Operator responsibility – Safety recommendations 8 2.2. EC Conformity The products described in the operating instructions conform to the requirement

Pagina 61

ME 9 2.4. Technical specifications Heating Immersion Circulator ME Working temperature range °C 20 ... 200 Temperature stability °C ±0,01