19510501-V2.doc 15.07.13 Operating Manual ME English Heating Immersion Circulator ME 1.951.0501-V2 06/13
Operator responsibility – Safety recommendations 10 Safety installations according to IEC 61010-2-010: Excess temperature protection adjustable fro
ME 11 Operating instructions 3. Safety notes for the user 3.1. Explanation of safety notes In addition to the safety warnings listed, warnings are
Safety notes for the user 12 3.3. Safety recommendations Follow the safety instructions to avoid personal injury and property damage. Also, the val
ME 13 Caution: The temperature controlling i.e. of fluids in a reactor constitutes normal circulator practice. We do not know which substances are
Operating controls and functional elements 14 4. Operating controls and functional elements 4.1. Circulator Front view Rear view 1 Mains p
ME 15 Menu keys 5 Key: start the menu > warning and safety values< 6 Key: start the menu >temperature setpoints< 7 Key: display of
Preparations 16 5. Preparations 5.1. Installation Caution: Securely fix the immersion circulator. The heater may not be in contact with the wall of t
ME 17 Recommended bath fluids: Bath fluid Temperature range soft/decalcified water 5 °C to 80 °C JULABO bath fluids JULABO Description T
Preparations 18 5.3. Temperature application to external systems Caution: Securely attach all tubing to prevent slipping. If the circulator is oper
ME 19 *) Adapter for metal tubing M10x1 on M16x1 Order No. 8 970 444 Warning: Tubing: At high working temperatures the tubing used for temp
2 Congratulations! You have made an excellent choice. JULABO thanks you for the trust you have placed in us. This operating manual has been design
Preparations 20 Pt100 RS 23215AMP15AMPVJULABO MVS2725a 25b26 For applications near ambient temperature (20 °C), the cooling water temperature shoul
ME 21 6. Operating procedures 6.1. Power connection Caution: Only connect the unit to a power socket with earthing contact (PE – protective earth)
Setting of temperatures 22 7. Setting of temperatures 7.1. Using the pre-settings in the menu Press the key to call up the menu for temperatu
ME 23 7.2. Direct setting of temperatures The circulator uses the setpoint of SETPNT 1 or 2 or 3 for temperature control. The indicated setpoi
Safety installations, warning functions 24 8. Safety installations, warning functions Check the safety installations at least twice a year! Refer
ME 25 8.1.1. Early warning system, low level protection This low level protection is independent of the control circuit and is divided into two
Safety installations, warning functions 26 8.2. Switch-over from warning to shutdown function If a shutdown of functional elements (e.g. heater, ci
ME 27 8.3. Over and Sub temperature warning function Over temperature Sub temperature If the observance of a working temperature value >SE
Menu functions 28 9. Menu functions Menu level 1 The term „Menu functions“ refers to settings such as Start program Page 29 Administration
ME 29 9.1. MENU PROGRAM – START Start-Menu This menu will start a previously set program. Requirements: 1. Create a program. (refer to next c
ME 3 TABLE OF CONTENTS Operating manual ...
Menu functions 30 Submenu TIMER Parameter level set the time for the start of the program in the submenu >TIMER<. >TIME<
ME 31 C2 D1 D2 E xxx.xx – external actual value D RUN – the program has started or PAUSE – the progress of the program has been interrupted by pr
Menu functions 32 Menu level 1 >EDIT< Create, administer program > STEP< Program step (1 ... 10) >SETPNT < Temperature
ME 33 9.3. MENU PUMP – Setting of pump pressure Factory setting: stage 2 The capacity of the circulating pump is set by adjusting the motor sp
Menu functions 34 9.4. MENU CONFIG – Configuration of unit Menu level 1 A RESET can be effected only in the >OFF< mode. Switch off c
ME 35 Level 3 Parameter level Year flashes, set by pressing + Return to factory settings by pressing RESET returns all set
Menu functions 36 interface. If such a safety standard is not required, the NAMUR recommendations can be bypassed with the AUTOSTART function thus all
ME 37 9.5. MENU CONTROL – Control characteristics and parameters Menu level 1 The circulator is qualified for internal and external
Menu functions 38 Level 2 Parameter level 0 … 999 The parameter flashes, set by pressing + 0.1 … 99.9 The parameter flashes,
ME 39 9.5.2. Dynamic internal °CtAPERNORMSetpointtemp. stability °CtAPERNORMtemperature ramptemp. stabilityThis parameter affects the temperature se
4 8.3. Over and Sub temperature warning function ... 27 9. Menu functions ...
Menu functions 40 9.5.4. Control parameters– XP-, TN-, TV- INTERNAL In most cases the control parameters preset in the factory are adequate for ach
ME 41 9.6. MENU SERIAL - BAUDRATE, HANDSHAKE, PARITY Menu level 1 For communication between circulator and a PC or a superordinated process con
Menu functions 42 9.7. MENU ATC - Absolut Temperature Calibration ATC serves to compensate a temperature difference that might occur between circul
ME 43 Menu level 1 Level 2 Parameter level Press the key if parameter is to be retained. Correction function for parameters or values (pri
Menu functions 44 The value is only indicated Integer digits flash, set by pressing + Decimal digits flash, set by pressing +
ME 45 9.7.3. CALIBRATION TYPE: 1 -/ 2 -/ 3 POINT A >1-point<, >2-point< or >3-point< calibration can be carried out. First geo
Menu functions 46 9.7.4. Example: 3-point calibration for internal control In the temperature range from 80 °C to 160 °C the calibration curve of
ME 47 The parameter flashes, switch by pressing and . A >3-point< calibration is carried out. The value >TMPVAL< is on
Menu functions 48 9.8. MENU LIMITS Menu level 1 Level 2 Parameter level Press the key if parameter is to be retained. Correction function
ME 49 9.8.1. Limits for internal control SETPOINT MAX / MIN – Maximum and minimum setpoint Restriction of the adjustable temperature range. Th
ME 5 Operating manual 1. Intended use JULABO circulators have been designed to control the temperature of specific fluids in a bath tank. JULABO
Menu functions 50 BAND HIGH / LOW – Band limitation The band limitation is active during external control. Varied, practice-oriented setting are fe
ME 51 10. Troubleshooting guide / error messages + Alarm with complete shutdown: If one of the following failures occur a complete, all-pole shut
Troubleshooting guide / error messages 52 Error message with ticker: >SENSOR DIFFERENCE ALARM-CHECK VISCOSITY AND PUMP STAGE< Defect of working
ME 53 11. Electrical connections Notice: Use shielded cables only. The shield of the connecting cable is electrically connected to the plug housing.
Remote control 54 / Control output The connector may be used for control of JULABO refrigerated circulators or as output for alarm messages. 1324
ME 55 12.2. Communication with a PC or a superordinated data system If the circulator is put into remote control mode via the configuration level,
Remote control 56 12.3. List of commands out commands: Setting temperature values or parameters. Command Parameter Response of circulator out_mode
ME 57 in commands: Asking for parameters or temperature values to be displayed. Command Parameter Response of circulator version none Number of s
Remote control 58 Command Parameter Response of circulator in_mode_01 none Selected setpoint: 0 = SETPNT 1 1 = SETPNT 2 2 = SETPNT 3 3 = Last setpo
ME 59 Error messages Description -06 SENSOR DIFFERENCE ALARM Sensor difference alarm. Working temperature and safety sensors report a temperature dif
Operator responsibility – Safety recommendations 6 2. Operator responsibility – Safety recommendations The products of JULABO ensure safe operation
JULABO Service – Online remote diagnosis 60 13. JULABO Service – Online remote diagnosis JULABO circulators of the HighTech series are equipped w
ME 61 14. Cleaning / repairing the unit Caution: Always turn off the unit and disconnect the mains cable from the power source before cleaning
ME 7 Use: The bath can be filled with flammable materials. Fire hazard! There might be chemical dangers depending on the bath medium used. Observe all
Operator responsibility – Safety recommendations 8 2.2. EC Conformity The products described in the operating instructions conform to the requirement
ME 9 2.4. Technical specifications Heating Immersion Circulator ME Working temperature range °C 20 ... 200 Temperature stability °C ±0,01